TM 5-3895-374-24-2
955-257
6.4
PART NUMBER CHANGES
The following list defines allowable part number changes for the part number table shown above.
DIGIT (1) : Any number change requires a hardware change too.
DIGIT (2) : Numbers 1, 2, 3, 4 are interchangeable by keypad instruction.
Numbers 5 and 6 are interchangeable by keypad instruction.
Numbers 7 and 8 are interchangeable by keypad instruction.
Any other change requires a hardware change too.
DIGIT (3) : Numbers 1, 2, 4 are interchangeable by keypad instructions only.
Any other changes require a hardware change too.
DIGIT (4) : Numbers OD thru 12 are interchangeable by keypad instructions.
Any other changes require a hardware change too.
DIGIT (5) : All numbers except 0 are interchangeable by keypad instruction.
DIGIT (6) : All numbers except 0 are interchangeable by keypad instruction.
DIGIT (7) : This number is used to identify specials and should only be changed by
authorized personnel.
Most recalibration requirements will use the original part number found on the instrument labels. If you make a change in
the PART NUMBER, be sure you write the new number down on the back cover of this manual for future reference. If you
do not need to reset or change the part number push, the PARAM DISPLAY key to advance the display and proceed to
paragraph 6.6.
If you do want to change the part number, proceed with paragraph 6.5.