These points are computed by the indicator whenever a
linearity correction is made. Re-entering these values will
restore the previous calibration.
98.
Channel Enable. (1)
Set to 1 to enable the channel.
Set to 0 to disable the channel.
Note:
You cannot disable channel 1. If you have not
connected scales to channels 2 or 3, make sure
that these channels are disabled.
99.
Exit Calibration Mode.
Selecting this function will exit the calibration mode and
return the indicator to the normal operating mode.
SECTION V - ERROR MESSAGES
1
Invalid Tolerance Selection.
2
Graduation Size Invalid.
3
Decimal Point Position Invalid (t - 4).
4
Flag Select must be 0 or 1.
5
Pushbutton Zero % must be 0 - 99.
6
Auto Zero % must be 0 - 99.
7
Only value 1 will reset parameters.
8
Only Value I will Reset Span Table.
9
Span exceeds Maximum Capacity and/or Entry too Small.
10
Cannot Span a Negative Value.
12
Motion Delay Time out of Range ( I - 255).
13
Starting Unit Value only 0 = LB, 1 · KG,.
14
Average Shift Value out of Range (1 = 1-50, 2 = 1-5000, 3 = 1-10000).
15
Overweight Entry out of Range (200 - 799.999).
16
Motion window out of Range (2 - 7999,999).
17
Increment Span Table Commands are 0 = No Incr., 1 = Incr.
18
Decrement Span Table Commands are 0 = No Decr., 1 Decr.
19
Span Table cannot be decremented past 1.
20
Parameter Memory Write Error. Indicator requires service.
21
Parameter Checksum Error. (Parameters have been lost.)
22
Program Check Fault. Indicator requires service.
23
Serial Port Speed Setting Range is 1 - 3.
24
Serial Port Mode Setting Range is I - 7.
25
Cannot Increment Span Table past 2
26
Entered offset larger than capacity