27
U.S.A. Display Flag is 0 = Off, 1 = On.
28
Power On Reset Warning is 0 = Off, 1 = On.
29
Channel Enable is 0 - Off, 1 = On
30
Only Value 1 will Set Deadload.
31
Average Register Select is 1, 2, 5 or 10.
32
Hold Off Delay Range is 0 - 30.
34
Load Cell Input Value out of Range (B - 35 mv).
35
Invalid Load Cell Value found in Table. Reload this Parameter.
37
Channel 1 cannot be turned Off.
38
Invalid function selection
40
Channel not Active.
41
Pushbutton Tare is Invalid (over/Under/Motion)
42
Keyboard Tare to one channel only.
43
Tare greater than Capacity.
44
Cannot do Dead Load Calculation, Input Signal is greater than
Parameter 11.
45
Cannot do Dead Load Calculation, Input Signal is too far Negative.
SECTION VI - TROUBLE SHOOTING GUIDE
The following is a list of problems you may encounter when installing or
operating your indicator.
Symptoms
Possible Cause
The indicator will not
- Check fuse in power supply
turn on - no display
- Check for AC power to supply and all
power connections.
Display shows uuuuuu
- Load cells not wired correctly (see
Section II)
- Defective load cell
- Dead load has not been set (see
Section III)
Display shows EEEEEE
- Scale is overloaded
- Overweight setting is not correct
(see Parameter 4)
- Dead load or Span not set properly
(see Section III)
- Load Cells not wired correctly (see
Section III
- Defective load cell
- Unused scale channels not disabled
(see Parameter 98)
Display shows rrrrrr.
- Dead load has not been set (see
Cannot zero display
Section III)
- Weight on scale is beyond Zero
Window (see Parameter 21)
- Scale has not been calibrated (see
Section III)
Display Unstable
- Defective load cell
- Load cell cabling wet or damaged
- Parameter 11 not set correctly (see
Section III)
- Defective Power supply
No output to Printer.
- Output connector not wired correctly
Setpoint not working
(see Section II)
- Output port not set-up correctly
(see Appendix I)
(page 3-71)